Skip to main content

Table 1 Parameters of system suitability for hydrocortisone and clotrimazole

From: Analytical method development and validation of hydrocortisone and clotrimazole in topical dosage form using RP-HPLC

Drugs

Retention time (min)

Instrument precision (RSD ≤ 2%)

Theoretical plates (≥ 2000)

Tailing factor (0.8–2.0)

Hydrocortisone

3.027

0.16

23462

1.177

3.025

0.29

23457

1.188

3.035

0.17

23548

1.188

3.046

0.36

23478

1.185

3.021

0.35

23386

1.185

3.015

0.22

23478

1.183

Clotrimazole

7.314

0.22

5298

1.892

7.303

0.42

5326

1.885

7.302

0.27

5308

1.896

7.368

0.23

5297

1.880

7.319

0.15

5326

1.890

7.313

0.10

5303

1.890