Skip to main content

Table 2 Validation parameters for hydrocortisone and clotrimazole assay

From: Analytical method development and validation of hydrocortisone and clotrimazole in topical dosage form using RP-HPLC

Working standards

Validation parameters

Linearity regression plot

Repeatability

Range (μg/ml)

Accuracy

LOQ (μg/ml)

LOD (μg/ml)

R

RSD (%)

Mean recovery

RSD (%)

Hydrocortisone

y = 15902x – 36428

0.9996

0.36

100–300

101.07

0.03

43.62

14.39

Clotrimazole

y = 60335x − 3142

0.9994

0.23

100–300

101.13

0.08

34.03

11.23