Skip to main content

Table 2 Parameters used for HPTLC

From: HPTLC analysis of Fumaria parviflora (Lam.) methanolic extract of whole plant

Parameters

Values

Calibration parameters

 Calibration mode

Single level

 Statistics mode

CV

 Evaluation mode

Peak height

Linomat 5 application parameters

 Spray gas

Inert gas

 Sample solvent type

Methanol

 Dosage speed

150 nl/s

 Predosage volume

0.2 μl

 Syringe size

100 μl

 Application position

8.0 mm

 Band length

8.0 mm

 Solvent front position

75.0 mm

Detection—CAMAG TLC scanner

 Number of tracks

2

 Position of track X

15.0 mm

 Distance between tracks

20.0 mm

 Scan start position Y

5.0 mm

 Scan end position Y

75. 0 mm

 Slit dimensions

6.00 × 0.30 mm, micro

 Optimize optical system

Light

 Scanning speed

20 mm/s

 Data resolution

100 μm/step

Integration: properties

 Data filtering

Savitsky-Golay 7

 Baseline correction

Lowest slope

 Peak threshold min. slope

5

 Peak threshold min. height

10 AU

 Peak threshold min. area

50

 Peak threshold max. height

990 AU

 Track start position

5.0 mm

 Track end position

75.0 mm

 Display scaling

Automatic

Measurement

 Wavelength

254 nm and 366 nm

 Lamp

D2/Hg

 Measurement type

Remission

 Measurement mode

Absorption/fluorescence

 Optical filter

Second order/K400

 Detector mode

Automatic

 PM high voltage

181 V