Table 1 Conclusive data with peak areas and standard deviation for system suitability

From: Method validation for bifenthrin emulsifiable concentrate and uncertainty calculation using gas chromatographic approach

Sr. #

Peak area of ISS solution

Peak areas of active

Area ratio

Retention time (min)

Standard deviation

RSD

Theoretical plates

ISS

Active

1

1823681

1834648

1.006

8.68

0.0027

0.27%

17999.012

17890.002

2

1824790

1825198

1.0002

17999.012

17890.002

3

1824790

1833184

1.0046

17870.674

17580.582

4

1824720

1832931

1.0045

17333.248

17990.111

5

1824854

1825212

1.0002

17970.624

17480.597