From: Determination of residual solvents in paclitaxel by headspace gas chromatography
Parameter | System conditions | Percentage RSD for peak area (n = 6) | Retention time | Plate count | Tailing factor |
---|---|---|---|---|---|
Flow rate (± 0.2 mL/min) | 2.3 | 2.2 | 9.07 | 31248 | 1.2 |
2.5 | 2.0 | 8.25 | 32548 | 1.0 | |
2.7 | 2.3 | 7.43 | 32478 | 1.3 | |
Column oven temperature (± 5 °C) | 35 °C | 2.1 | 9.08 | 30214 | 1.2 |
40 °C | 2.0 | 8.25 | 32548 | 1.0 | |
45 °C | 2.3 | 7.45 | 31067 | 1.4 | |
Vial oven temperature (± 5 °C) | 85 °C | 2.1 | 8.24 | 30148 | 1.3 |
90 °C | 2.0 | 8.25 | 32548 | 1.0 | |
95 °C | 1.9 | 8.21 | 31032 | 1.1 |