Skip to main content

Table 5 Intra and inter assay precision data (n = 6)

From: Development and validation by statistical treatment of stability indicating RP-HPLC method for quantification of Orlistat in Orlistat-loaded solid dispersion

Actual concentration (μg/ml) Measured concentration (μg/ml), ± RSD (%)
Intra day Inter day
2.5 2.48, 2.02 2.61, 2.86
5 5.1, 1.64 5.77, 2.82
7.5 7.48, 1.78 7.73, 2.69