Skip to main content

Table 5 Accuracy study

From: HPLC method validation for quantification of tetrahydrocurcumin in bulk drug and formulation

Level

% Recovery

Mean % recovery

%RSD

50%

98.24

98.23

0.194

 

98.42

 

98.04

100%

98.31

98.32

0.016

 

98.32

 

98.34

150%

100.01

99.99

0.021

 

99.97

 

100.00