From: HPLC method validation for quantification of tetrahydrocurcumin in bulk drug and formulation
Parameter | RT | Asymmetry factor | Theoretical plates | % Assay | |
---|---|---|---|---|---|
Change in wavelength | +3 nm | 6.61 | 1.15 | 5671 | 98.56 |
−3 nm | 6.62 | 1.17 | 5661 | 100.36 | |
Flow rate | +10% ml/min | 6.64 | 1.18 | 5680 | 100.17 |
−10% ml/min | 6.66 | 1.18 | 5652 | 99.09 | |
Column temperature | +2 °C | 6.56 | 1.16 | 5627 | 100.38 |
−2 °C | 6.56 | 1.16 | 5641 | 100.33 |