Skip to main content

Table 1 Summary of forced degradation study results

From: Development of new stability indicating UPLC-UV method for the extraction and quantification of perindopril and indapamide from human plasma

Stress condition

Time

Assay% of PP after degradation

Assay% of IP after degradation

Remarks

Unstressed sample

–

99.63

100.2

–

Acid hydrolysis (1.0N HCl)

60 min

91.15

98.64

The significant degradation was observed.

Base hydrolysis (1.0N NaOH)

60 min

96.01

93.78

The slight degradation of analytes was observed.

Oxidation (10% H2O2)

60 min

98.65

99.12

Significant degradation was not observed.

Thermal (80 ° C)

8 h

99.32

99.81

Significant degradation was not observed.

Photolytic degradation

24 h

99.53

99.98

Significant degradation was not observed.