Skip to main content

Table 3 Precision table of Ivacaftor in mobile phase and human plasma

From: A rapid RP-HPLC method for the simultaneous estimation of Ivacaftor and Tezacaftor and in silico study of their metabolitic products

S. no.

Conc (μg/mL)

Mobile phase

Human plasma

Conc found

(μg/mL)

%

Avg %

Conc found

(μg/mL)

%

Avg %

1

45

44.99

99.9

100.06

45.95

102.1

100.02

2

44.78

99.5

44.71

99.35

3

44.99

99.9

45.07

100.1

4

44.88

99.7

43.76

97.24

5

45.29

100.6

45.05

100.1

6

45.39

100.8

45.53

101.1

S.D

 

0.096

 

0.484

% R.S.D

 

0.21

 

1.07