Skip to main content

Table 5 Intermediate precision of Ivacaftor in mobile phase and human plasma

From: A rapid RP-HPLC method for the simultaneous estimation of Ivacaftor and Tezacaftor and in silico study of their metabolitic products

S. no.

Conc (μg/mL)

Mobile phase

Human plasma

Conc found

(μg/mL)

%

Avg %

Conc found

(μg/mL)

%

Avg%

1

45

46.53

103.4

99.15

45.94

102.0

101.7

2

44.6

99.1

46.01

102.2

3

44.26

98.3

45.68

101.5

4

44.54

98.9

45.71

101.5

5

43.84

97.4

45.72

101.6

6

44.04

97.8

45.68

101.5

S.D

 

0.396

0.13

% R.S.D

 

0.8

0.28