Parameter | Conditions |
---|---|
Eluent | 0.1% Formic acid in water/acetonitrile—50/50 (v/v) |
Flow rate | 1.0 mL min−1 |
Auto-sampler temperature | 25 °C |
Injection volume | 20 µL |
Column temperature | 30 °C |
Elution | Isocratic |
Run time | 10 min |
Source | ESI |
Ionization mode | Positive |
Capillary voltage | 5 kV |
Cone voltage | 30 V |
Acquisition mode | MRM |
Parent m/z selection for quantification | 125.1 Da |
Source temperature | 150 °C |
Desolvation temperature | 320 °C |
Drying and nebulization gas | Nitrogen |
Nebulization gas flow | 1.5 mL min−1 |
Drying gas flow | 15 mL min−1 |
CID gas | Argon |
Collision energy | 40 V |