Skip to main content

Table 1 Summary of validation data

From: Stability-indicating RP-HPLC method development and validation for estimation of Mupirocin calcium in bulk and in pharmaceutical formulation

Linear range [µg/mL] [n = 5] 4–24 µg/mL
Correlation coefficient 0.9998
Limit of detection [µg/mL] 0.35
Limit of quantification [µg/mL] 1.08
% Recovery [n = 3] 99.41–99.80
Precision [%RSD]  
Repeatability [n = 6] 0.27
Intra-day [n = 3] 0.28–0.61
Inter-day [n = 3] 0.04–0.25
Ruggedness [% RSD]  
Analyst – I [n = 6] 0.54
Analyst – II [n = 6] 0.51
Robustness Robust