Skip to main content

Table 9: Summary of validation parameters

From: Simultaneous estimation for Dicyclomine HCl and Simethicone in bulk and oral liquid drop formulation: an RP-HPLC method development and validation

Parameters

Dicyclomine HCl

Simethicone

Linearity range (ng/ml)

5–40 μg/ml

5–40 μg/ml

Regression equation

Y = 0.1494X + 0.0007

Y = 0.7418X − 0.0039

Correlation Coefficient (r2)

0.9995

0.9991

Accuracy (% RSD)

0.95

1.30

Precision (% RSD)

0.42

0.45

Repeatability (n = 5)

  

 Intra-day (n = 3)

0.48

0.50

 Inter-day (n = 3)

0.49

0.49

Robustness

  

 Analyst I (n = 6)

0.45

0.49

 Analyst II (n = 6)

0.49

0.51

LOD (μg/ml)

0.72

0.57

LOQ (μg/ml)

2.19

1.73