Skip to main content

Table 6 Results of accuracy study

From: A new validated stability-indicating RP-HPLC method for simultaneous quantification of dolutegravir and lamivudine in bulk and pharmaceutical dosage form

% of target concentration Lamivudine (% recovery) Dolutegravir (% recovery)
50 100.16 101.48
100 99.03 99.21
150 100.21 100.48
Mean % recovery (n = 3) 99.8 100.4
%RSD 0.67 1.13
  1. At each % level mean % recovery in the acceptable limit of 98.0 to 102.0%