Trial | Column | Mobile phase | Observation |
---|---|---|---|
1 | BDS C18 (250 mm × 4.6 mm, 5 μm) | Methanol: 0.1 % OPA (40:60) | Only MET was eluted with less theoretical plates |
2 | BDS C18 (250 mm × 4.6 mm, 5 μm) | ACN: KH2PO4 (40:60) | Only MET was eluted |
3 | Agilent C18 (250 mm × 4.6 mm, 5 μm) | ACN: 0.1 % OPA (60:40) | Peak symmetry of both the peaks was good but with increased retention time |
4 | Kromasil C18 (150 mm × 4.6 mm, 5 μm) | ACN: 0.1 % OPA (50:50) | Peak symmetry was good but MET eluted at 1.6 min |