Skip to main content

Table 1 System suitability study

From: HPLC method validation for quantification of tetrahydrocurcumin in bulk drug and formulation

Injection

Retention time

Peak area

Asymmetry factor

Theoretical plates

1

6.62

19186889

1.16

5943

2

6.62

19126526

1.15

5997

3

6.62

19234275

1.15

5985

4

6.62

19198526

1.15

5973

5

6.63

19183676

1.15

5956

6

6.63

19196342

1.16

5883

Mean

6.623

19187705.67

1.153

5956.17

SD

0.005

34986.36

0.005

40.76

%RSD

0.078

0.18

0.448

0.68