Skip to main content

Table 7 Precision studies for standard and test solutions of sofosbuvir

From: Development and validation of RP HPLC method for the estimation of Sofosbuvir and related impurity in bulk and pharmaceutical dosage form

S. no.

Retention time (min)

Peak area

Standard

Test

Standard

Test

1

3.756

3.757

6012506

5971170

2

3.761

3.755

5972456

5972988

3

3.763

3.755

5972404

5969403

4

3.832

3.755

6231754

5967112

5

3.756

3.757

5967803

5973215

6

3.755

3.755

5967236

5973758

Mean

3.771

3.756

6020693

5971274

% RSD

0.807

0.026

1.741

0.043

SD

0.030

0.001

104812

2597