Sample | Area | % Absolute difference | Acceptance criteria | Conclusion |
---|---|---|---|---|
Unfiltered | 16,256,478 | NA | % Absolute difference NMT 2.0 | Both PVDF and Nylon filters passes the criteria for filter study |
0.45 μ PVDF filter | 16,178,521 | 0.48 | ||
0.45 μ Nylon filter | 16,152,546 | 0.64 |