Skip to main content

Table 2 System suitability parameters

From: Analytical method validation for related substances in sodium valproate oral solution by gas chromatography

System suitability parameters

Observed value

Acceptance criteria

The % RSD of six replicate standard injections

1.3

NMT 2.0%

The tailing factor for the Sodium Valproate peak from the first standard injection

1.3

NMT 2.0

The Theoretical plate for the Sodium Valproate peak from the first standard injection

64,907

NLT 2000

  1. NMT Not more than, NLT not less than