Skip to main content

Table 6 Accuracy/recovery studies for sodium valproate

From: Analytical method validation for related substances in sodium valproate oral solution by gas chromatography

Recovery level

Sample no.

‘ppm’ added

‘ppm’ recovered

% Recovery

LOQ %

1

0.5044

0.5057

100.25

2

0.5044

0.5119

101.48

3

0.5044

0.5049

100.09

100%

1

1.0089

1.0091

100.02

2

1.0089

1.0011

99.23

3

1.0089

1.0020

99.32

150%

1

1.5133

1.4821

97.94

2

1.5133

1.4877

98.31

3

1.5133

1.4803

97.82

250%

1

2.5222

2.3883

94.69

2

2.5222

2.3515

93.23

3

2.5222

2.3285

92.32

 

Average

97.89

% RSD

2.00