Skip to main content

Table 4 Precision results for triamterene and hydrochlorothiazide

From: Development and validation of stability indicating UPLC method for the simultaneous estimation of triamterene and hydrochlorothiazide in combined dosage forms using quality by design approach

Injection

Triamterene area

HCTZ area

1

838,408

245,103

2

835,617

246,003

3

832,556

240,386

4

837,966

246,088

5

830,038

249,300

6

839,801

245,551

Ave

835,731.0

245,405.2

SD

3774.4

2874.0

Percent RSD

0.5

1.2

  1. Acceptance: The percent relative standard deviation should not be more than 2%