Table 4 Robustness results of the proposed method

From: Development and validation of a stability-indicating RP-HPLC method for estimation of Lefamulin in pure and pharmaceutical drug products

Chromatographic conditions

Deliberate changes

Retention time

Tailing factor

Theoretical plates count

% RSD of peak areaa

Flow rate (mL/min)

0.9

3.95

1.09

4705

0.35

1.1

2.40

1.39

5450

0.21

Wavelength (nm)

208 nm

3.08

1.12

5204

0.19

212 nm

3.09

1.13

5310

0.42

Column temperature (°C)

27

3.21

1.09

5023

1.04

33

2.94

1.18

5125

0.49

Acetonitrile ratio

48.5%

3.47

1.14

4941

0.58

51.5%

2.89

1.23

4532

0.79

Acceptance value

0.8–2.0

> 2000

< 2

  1. aMean of three measurements