Skip to main content

Table 6 System suitability parameters

From: New analytical LC–MS/MS method for fluconazole and ivermectin estimation in combined pharmaceutical dosage form: development and validation

Parameters

Fluconazole

Ivermectin

Flow rate (mL/min)

0.400

0.400

Retention time (min)

1.10

1.05

Peak area

46,760.69

6442.63

Trace

307 > 238

897 > 897

Linearity (ppb)

100–1250

100–1250

MRM (M+)

169

897

LOD (g/mL)

0.0034

0.010

LOQ (g/mL)

0.074

0.225