Skip to main content

Table 10 Table for method precision

From: Development and validation of ICPMS methods for simultaneous determination of elemental impurities in topical cream containing ximenynic acid

Aspirations

Ratio of analyte response to internal standard response

V

Co

Ni

As

Cd

Hg

Pb

1

9.4547

9.3576

22.0298

0.5218

0.8007

0.0926

0.1835

2

10.5279

10.3086

23.6646

0.5681

0.8581

0.1029

0.2018

3

9.6006

9.3929

21.7001

0.5205

0.7853

0.0938

0.1852

4

9.5205

9.3604

21.7120

0.5287

0.7915

0.0931

0.1845

5

9.5716

9.3646

21.4316

0.5273

0.8017

0.0928

0.1846

6

9.7889

9.4586

21.6740

0.5227

0.8178

0.0941

0.1858

Mean

9.7440

9.5405

22.0354

0.5315

0.8092

0.0949

0.1876

SD

0.40010

0.37820

0.82060

0.01820

0.02640

0.00400

0.00700

%RSD

4.11

3.96

3.72

3.42

3.26

4.21

3.73