Skip to main content

Table 6 Summary of validation parameters

From: Development of a robust and reliable reverse-phase high-performance liquid chromatography (RP-HPLC) method using analytical quality by design principles for the accurate determination of esculin in its bulk form

Validation parameters

Results

Linearity

 Correlation coefficient

0.9998

 Range

4–20 µg/ml

Sensitivity

 LOD (µg/ml)

0.82891

 LOQ (µg/ml)

2.511

Precision

 Inter-day (%RSD)

0.665

 Intra-day (%RSD)

0.612

Robustness

0.9089

 Flow rate (%RSD)

0.7385

 Ratio (%RSD)

0.9365

 Temperature (%RSD)

1.0516

Ruggedness

 

 Change in analyst (%RSD)

1.1011

System suitability

1.24

 Retention time (%RSD)

0.1398

 Peak area (%RSD)

0.7146

 Theoretical plates (%RSD)

0.0612