Skip to main content

Table 4 Intra-day precision values at different levels of the validated analytical method

From: QbD-driven RP-HPLC method for novel chemo-herbal combination, in-silico, force degradation studies, and characterization of dual drug-loaded polymeric and lipidic nanocarriers

Analytes

Actual concentration (µg/mL)

Observed Concentration* (µg/mL)

Intra-day precision %RSD

% Accuracy*

Avg. % Bias

BCA

0.5

0.495 (± 0.006)

1.130

98.998 (± 1.12)

1.002

4

3.98 (± 0.044)

1.102

99.499 (± 1.1)

0.5015

32

31.812 (± 0.619)

1.947

99.413 (± 1.94)

0.5874

LTB

0.5

0.518 (± 0.007)

1.403

103.86 (± 1.455)

− 3.7763

4

3.954 (± 0.07)

1.927

98.86 (± 1.905)

1.1392

32

31.84 (± 0.573)

1.800

99.51 (± 1.791)

0.4894

  1. *Data expressed as mean ± SD, n = 3