Skip to main content

Table 3 Method validation parameters

From: A validated, precise TLC-densitometry method for simultaneous quantification of mahanimbine and koenimbine in marketed herbal formulations

Sl No

Parameters

Mahanimbine

Koenimbine

1

Wavelength

285 nm

291 nm

2

Linearity range concentration (in µg/ml)

100–400

50–450

3

Coefficient of variances (CV)

0.34%

0.89%

4

Correlation coefficients (R2)

0.9985

0.9998

5

Limit of detection (ng/spot)

32.81

18.44

6

Limit of quantitation (ng/spot)

72.81

31.57

7

Reproducibility

Reproducible

Reproducible