Skip to main content

Table 4 Robustness

From: A new gas chromatographic method for quantification of Metformin hydrochloride and Vildagliptin in bulk and pharmaceutical dosage form: development and validation

Parameter changes

Retention time (Min)

Peak Area

MET

VID

MET

VID

The initial temperature of the oven

98 °C

10.225

22.045

5,486,025

605,845

100 °C

10.204

22.021

5,475,845

603,654

102 °C

10.189

21.995

5,474,554

602,965

%RSD

0.1772

0.1136

0.1147

0.2489

Flow rate (mL/min)

0.9

10.225

22.052

5,485,445

605,868

1.0

10.205

22.018

5,475,841

603,674

1.1

10.155

22.001

5,467,828

601,964

%RSD

0.3537

0.1179

0.1611

0.3241